{"cve":{"cve_id":"CVE-2024-9832","is_kev":false,"kev_date_added":null,"kev_vendor_project":null,"kev_product":null,"kev_vulnerability_name":null,"kev_short_description":null,"kev_required_action":null,"kev_due_date":null,"kev_known_ransomware":null,"kev_notes":null,"kev_cwes":null,"epss_score":0.00177,"epss_percentile":0.0737,"epss_as_of":"2026-06-23","description":"There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.","published_at":"2024-11-14T21:03:16.721000Z","last_modified_at":null,"cvss_v3_score":9.3,"cvss_v3_vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H","cvss_v3_severity":"CRITICAL","cvss_v4_score":null,"cvss_v4_vector":null,"cvss_v4_severity":null,"ssvc_decision":null,"ssvc_exploitation":null,"ssvc_automatable":null,"ssvc_technical_impact":null,"cwes":["CWE-307"],"nvd_references":["https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"],"vuln_status":null,"trending_score":null,"is_trending":false,"has_trended":false,"trended_number_one":false,"trending_peak_score":null,"trending_peak_rank":null,"started_trending_at":null,"trended_number_one_at":null,"summary_generated":null,"summary_generated_at":null,"summary_model":null,"created_at":"2026-06-24T00:09:39.878444Z","updated_at":"2026-06-28T23:24:58.075290Z"},"effective_severity":"CRITICAL","badges":[],"impact_analysis":[],"cvss_v3_decoded":{"version":"3.1","metrics":[{"metric":"AV","name":"Attack Vector","value":"L","value_label":"Local"},{"metric":"AC","name":"Attack Complexity","value":"L","value_label":"Low"},{"metric":"PR","name":"Privileges Required","value":"N","value_label":"None"},{"metric":"UI","name":"User Interaction","value":"N","value_label":"None"},{"metric":"S","name":"Scope","value":"C","value_label":"Changed"},{"metric":"C","name":"Confidentiality","value":"H","value_label":"High"},{"metric":"I","name":"Integrity","value":"H","value_label":"High"},{"metric":"A","name":"Availability","value":"H","value_label":"High"}]},"cvss_v4_decoded":{"version":null,"metrics":[]},"affected":[{"vendor_slug":"baxter","vendor_name":"Baxter","product_slug":"life2000-ventilation-system","product_name":"Life2000 Ventilation System","version_start":"06.08.00.00 and prior","version_start_inclusive":true,"version_end":"06.08.00.00 and prior","version_end_inclusive":true,"cpe23_uri":"cve5:baxter:life2000-ventilation-system:06.08.00.00 and prior:06.08.00.00 and prior"}],"exploit_refs":[],"news":[],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01","source_type":"VENDOR_ADVISORY","tags":["advisory"]}],"timeline":[{"type":"published","at":"2024-11-14T21:03:16.721000Z","label":"CVE published","source":null},{"type":"cvss_changed","at":"2026-06-28T17:40:39.588970Z","label":"CVSS score revised","source":"cvelistv5"},{"type":"cvss_changed","at":"2026-06-28T17:40:39.588970Z","label":"CVSS score revised","source":"cvelistv5"},{"type":"cvss_changed","at":"2026-06-28T17:40:39.588970Z","label":"CVSS score revised","source":"cvelistv5"}]}